General Properties
* Laser scatter + Chemical dye + Flow cytometry
* 5-part differentiation, 29 parameters
* Low sample volume as 20 ul
* Up to 80samples per hourwith autoloader
* Loading up to 60 samples simultaneously
* Support continuous sample loading
* Peripheral blood, venous blood and pre-diluted blood counting
* Close and open vial modes
* STAT function
* Powerful capability to flag abnormal cellsand up to 20 user-defined alarm ranges
* Working status self-test and reagent volume self-detection
* Automated test re-running, with defined rules
* Error removal and sleep/wakeup functions
* Large storage capacity: up to 100,000 samples
* Recommended or customizable decision rules for re-exam abnormal samples
* QC and calibration management
* Support uni- or bi-directional LIS
Tecnical Specifications
Principles
Semiconductor laser scatter, Flow Cytometry (WBC Diff)
Impedance method for WBC, RBC, PLT counting
Cyanide-free reagent for Hemoglobin test
Parameters
WBC, Lym%, Lym#,Mon%, Mon#,Neu%, Neu#,Eos%, Eos#, Bas%, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PDW, MPV, PCT, P-LCC and, P-LCR and 4 research parameters LIC%, LIC#, ALY%, ALY#
1 3D and 3 2D 3 scatter grams and 3 histograms
Throughput
Up to 80 samples per hour with autoloader
Sample Volume
≤20ul
Test Mode
CBC CBC+DIFF
Counting Modes
Whole Blood, pre-diluted blood, peripheral blood
Performance
Carryover Reproducibility (CV%) Linearity
WBC ≤0.5% ≤2.0% (4-15×109/L) 0.00-300×109/L
RBC ≤0.5% ≤1.5% (3.5-6.0×1012/L) 0.00-8.5×1012/L
HGB ≤0.5% ≤1.5% (110-180g/L) 0-250g/L
PLT ≤1.0% ≤4.0% (150-500×109/L) 0.00-3000×109/L
Data Storage Capacity
Up to 100,000 results including numeric and graphical information
Communication
LAN port supports HL7 protocol
Operating Environment
Temperature: 15oC-30oC
Humidity: 30-85%
Air Pressure: 70-106 kPa
Power Requirement
A.C. 100-240V ≤300VA 50/60Hz
Dimension and Weight
620mm(L) × 620mm(W) × 535mm(H)
Weight: 56kg